10-10 DICOM
Name Description
[Verify]
Click to verify if the two DICOM application entities are normally
connected.
[Exit] Click to exit the screen.
Note:
In terms of !Scheduled Station AE Title", if you set this item in the Worklist server, then
!Scheduled Station AE Title" configured here should be consistent with the one set in the
server.
For example, when you set !Scheduled Station AE Title" in the Worklist server as AAA (this
means you assigned the scheduled exam to AAA), then both !Scheduled Station AE Title" and
!AE Title" in the ultrasound system configuration should be AAA.
10.1.3.4 Storage Commitment
1. Enter [Setup]-> [DICOM Preset]-> [Set DICOM Service]-> [Storage Commitment].
2. Select device, enter the information. For device setting, please refer to !10.1.3 DICOM
Service."
z Click [Add] to add the service to the Service list;
z Select an item in the service list, change the parameters in the above area, and click
[Update] to update the item in the service list; click [Cancel] to cancel changing.
z Select an item in the service list, and click [Delete] to delete the service.
z Select an item in the service list, and click [Default] to set the server to be the default
service.
3. Select an item in the service list, and click [Verify] to verify the connection.
4. Click [Exit] to confirm the setting and exit the page.
DICOM storage commitment setting items are described as follows:

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer MAGIC 6000 and is the answer not in the manual?

Eickemeyer MAGIC 6000 Specifications

General IconGeneral
BrandEickemeyer
ModelMAGIC 6000
CategoryMedical Equipment
LanguageEnglish

Related product manuals