Acoustic Output Reporting Table (60601-2-37) B-11
Transducer Model: 65C15EA
Imaging Mode: B-mode
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1 cm
2
A
aprt
> 1
cm
2
Maximum Index Value 6.25E-01 2.45E-01 2.98E-01
Associate
Acoustic
Parameter
p
r.a
1.39E+00
P 1.04E+01 1.04E+01
Min of [P
Į
(z
s
), I
ta,Į
(z
s
)]
z
s
z
bp
z
b
z at max.I
pi,Į
1.75E+00
d
eq
(z
b
)
f
awf
5.06E+00 5.09E+00 5.06E+00
Dim of A
aprt
X
1.02E+00 1.02E+00
Y 6.00E-01 6.00E-01
Other
Informatio
n
t
d
3.66E-01
prr 5.61E+03
p
r
at max. I
pi
1.88E+00
d
eq
at max. I
pi
I
pa,Į
at max. MI 8.14E+01
Focal
Length
FL
x
(cm)
8.00E+00 9.00E+00
FL
y
(cm)
3.50E+00 3.50E+00
Operating
Control
Conditions
Focus position
3cm 8cm 8cm 9cm
Display depth
12cm 12cm 12cm 12cm
Acoustic power
100% 100% 100% 100%
Focus number one one one one
Working Frequency H9M H9M H9M 6.5M
Scan Line Mode
Lower-density line and minimum scan angle