B-12 Acoustic Output Reporting Table (60601-2-37)
Transducer Model: 65C15EA
Imaging Mode: B+M-mode
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-
scan
A
aprt
1 cm
2
A
aprt
> 1
cm
2
Maximum Index Value 6.70E-01 3.50E-01 3.04E-01 3.35E-01
Associate
Acoustic
Parameter
p
r.a
1.33E+00
P 2.01E+00 2.01E+00 2.01E+00
Min of [P
Į
(z
s
), I
ta,Į
(z
s
)]
z
s
z
bp
z
b
1.84E+00
z at max.I
pi,Į
1.95E+00
d
eq
(z
b
) 2.01E-01
f
awf
5.08E+00 5.21E+00 5.05E+00
Dim of A
aprt
X
9.55E+00 1.02E+00
Y 3.68E+00 6.00E-01
Other
Informatio
n
t
d
3.56E-01
prr 2.00E+03
p
r
at max. I
pi
2.10E+00
d
eq
at max. I
pi
2.01E-01
I
pa,Į
at max. MI 1.26E+02
Focal
Length
FL
x
(cm)
8.00E+00 8.00E+00
FL
y
(cm) 3.50E+00 3.50E+00
Operating
Control
Conditions
Focus position
3cm 8cm 8cm 9cm
Display depth
12cm 12cm 12cm 12cm
Acoustic power
100% 100% 100% 100%
PRF 2000 2000 2000 2000
Working Frequency H9M H9M H9M 6.5M
Scan Line Mode Lower-density line and minimum scan angle