Cerabar M PMC51, PMP51, PMP55
Endress+Hauser 11
PMP55
The operating pressure acts on the process isolating diaphragm of the diaphragm seal and is
transferred to the process isolating diaphragm of the sensor by a diaphragm seal fill fluid. The
process isolating diaphragm is deflected and a fill fluid transfers the pressure to a resistance bridge.
The pressure-dependent change in the bridge output voltage is measured and evaluated.
Advantages:
Depending on the version, can be used for process pressure up to 400 bar (6 000 psi) and
simultaneous extreme process temperatures
High long-term stability
Guaranteed overload resistance up to 4 times the nominal pressure
Level measurement (level,
volume and mass)
h =
p
ρ g
h
A0023678
h Height (level)
p Pressure
ρ Density of the medium
g Gravitation constant
Your benefits
Choice of different level measuring modes in the device software
Volume and mass measurements in any tank shapes by means of a freely programmable
characteristic curve
Choice of diverse level units
Has a wide range of uses, even in the following cases:
in the event of foam formation
in tanks with agitators of screen fittings
in the event of liquid gases
Electrical differential
pressure measurement with
gauge pressure sensors
Fieldgate
FXA520
FXN 520
FXN 520
Multidrop-Connector
FXN520
Cerabar M
Á
11
2
A0023680
1 Shut-off valves
2 e.g. filter
In the example given, two Cerabar M devices (each with a gauge pressure sensor) are
interconnected. The pressure difference can thus be measured using two independent Cerabar M
devices.