EasyManua.ls Logo

Endress+Hauser Cerabar M PMC51 - PMP55: Process Connections with Flush-Mounted Process Isolating Diaphragm

Endress+Hauser Cerabar M PMC51
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar M PMC51, PMP51, PMP55
82 Endress+Hauser
PMP55: process connections
with flush-mounted process
isolating diaphragm
Pipe diaphragm seal Tri-Clamp ISO 2852
d
2
d
3
h
L
D
A0021671
h Height
D Diameter
d
2
Diameter
d
3
Diameter
L Installed length
Engineering unit mm (in)
Material
1
)
Nominal
diameter
ISO 2852
Nominal
diameter
Nominal
pressure
D d
2
d
3
h L Weight Approval
2)
Option
3)
[in] [mm] [mm] [mm] [mm] [mm] [kg (lb)]
AISI 316L
DN 10 ¾ PN 40 10.5 25 34 41.5 140 0.6 (1.32) 3A, CRN SIJ
DN 25 1 PN 40 22.5 50.5 54 67 126 1.7 (3.75) 3A, CRN SBJ
DN 38 1 ½ PN 40 35.5 50.5 69 67 126 1.0 (2.21) 3A, CRN SCJ
4)
DN 51 2 PN 40 48.6 64 78 79 100 1.7 (3.75) 3A, CRN SDJ
4)
1) Surface roughness of the surfaces in contact with the medium R
a
< 0.76 µm (29.9 µin) as standard.
2) CSA approval: Product Configurator, order code for "Approval"
3) Product Configurator, order code for "Process connection"
4) incl. 3.1 and pressure test according to Pressure Equipment Directive, Category II

Table of Contents