Cerabar M PMC51, PMP51, PMP55
Endress+Hauser 33
Reference accuracy [E1]
Reference accuracy comprises the non-linearity [IEC 62828-1 / DIN EN 61298-2] including the
hysteresis [IEC 62828-1 / DIN EN 61298-2] and the non-repeatability [IEC 62828-1 / DIN EN
61298-2] in accordance with the limit point method as per [IEC 62828-1 / DIN EN 60770-2].
PMP51
400 mbar (6 psi) sensor
• Standard: TD 1:1 = ±0.1 %; TD >1:1 to 20:1 = ±0.15 % · TD
• Platinum: -
1 bar (15 psi) sensor
• Standard: TD 1:1 to 5:1 = ±0.1 %; TD >5:1 to 20:1 =±0.03 % · TD
• Platinum: TD 1:1 to 2.5:1 = ±0.075 %; TD >2.5:1 to 20:1 =±0.03 · TD
2 bar (30 psi) sensor
• Standard: TD 1:1 to 10:1 = ±0.1 %; TD >10:1 to 20:1 = ±0.015 % · TD
• Platinum: TD 1:1 to 5:1 = ±0.075 %; TD >5:1 to 20:1 = ±0.015 % · TD
4 bar (60 psi) sensor
• Standard: TD 1:1 to 10:1 = ±0.1 %; TD >10:1 to 20:1 = ±0.2 %
• Platinum: TD 1:1 to 10:1 = ±0.075 %; TD >10:1 to 20:1 = ±0.0075 % · TD
10 bar (150 psi) and 40 bar (600 psi) sensor
• Standard: TD 1:1 to 10:1 = ±0.1 %; TD >10:1 to 20:1 = ±0.2 %
• Platinum: TD 1:1 to 10:1 = ±0.075 %; TD >10:1 to 20:1 = ±0.1 %
100 bar (1 500 psi) sensor
• Standard: TD 1:1 to 10:1 = ±0.1 %; TD >10:1 to 20:1 = ±0.2 %
• Platinum: TD 1:1 to 10:1 = ±0.075 %; TD >10:1 to 20:1 = ±0.15 %
400 bar (6 000 psi) sensor
• Standard: TD 1:1 to 5:1 = ±0.1 %; TD >5:1 to 20:1 = ±0.03 % · TD
• Platinum: TD 1:1 to 5:1 = ±0.1 %; TD >5:1 to 20:1 = ±0.03 % · TD
PMP51 with hygienic process connections
400 mbar (6 psi) sensor
• Standard: TD 1:1 = ±0.1 %; TD >1:1 to 10:1 = ±0.3 % · TD
• Platinum: -
1 bar (15 psi) sensor
• Standard: TD 1:1 = ±0.1 %; TD >1:1 to 10:1 = ±0.3 % · TD
• Platinum: TD 1:1 = ±0.1 %; TD >1:1 to 10:1 = ±0.2 % · TD
2 bar (30 psi) sensor
• Standard: TD 1:1 to 5:1 = ±0.1 %; TD >5:1 to 10:1 = ±0.2 %
• Platinum: TD 1:1 to 5:1 = ±0.075 %; TD >5:1 to 10:1 = ±0.1 %
4 bar (60 psi), 10 bar (150 psi) and 40 bar (600 psi) sensor
• Standard: TD 1:1 to 10:1 = ±0.1 %; TD >10:1 to 20:1 = ±0.2 %
• Platinum: TD 1:1 to 10:1 = ±0.075 %; TD >10:1 to 20:1 = ±0.1 %
PMP55
400 mbar (6 psi) Sensor
• Standard: TD 1:1 = ±0.15 %; TD >1:1 to 20:1 = ±0.15 % · TD
• Platinum: -
1 bar (15 psi) Sensor
• Standard: TD 1:1 to 5:1 = ±0.15 %; TD >5:1 to 20:1 = ±0.03 % · TD
• Platinum: TD 1:1 to 2.5:1 = ±0.075 %; TD >2.5:1 to 20:1 = ±0.03 % · TD
2 bar (30 psi) Sensor
• Standard: TD 1:1 to 10:1 = ±0.15 %; TD >10:1 to 20:1 = ±0.015 % · TD
• Platinum: TD 1:1 to 5:1 = ±0.075 %; TD >5:1 to 20:1 = ±0.015 % · TD
4 bar (60 psi) Sensor
• Standard: TD 1:1 to 10:1 = ±0.15 %; TD >10:1 to 20:1 = ±0.2 %
• Platinum: TD 1:1 to 10:1 = ±0.075 %; TD >10:1 to 20:1 = ±0.0075 % · TD
10 bar (150 psi) and 40 bar (600 psi) Sensor
• Standard: TD 1:1 to 10:1 = ±0.15 %; TD >10:1 to 20:1 = ±0.2 %
• Platinum: TD 1:1 to 10:1 = ±0.075 %; TD >10:1 to 20:1 = ±0.1 %