EasyManua.ls Logo

Endress+Hauser Cerabar M PMC51 - Installation; Incoming acceptance and storage; Installation conditions

Endress+Hauser Cerabar M PMC51
134 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar M PMC51, PMP51, PMP55
Endress+Hauser 9
OPL
PMC51: max. 60 bar (900 psi)
PMP51: max. 600 bar (9 000 psi)
PMP55: max. 600 bar (9 000 psi)
Process temperature range
PMC51: –20 to +130 °C (–4 to +266 °F)
For a maximum of 60 minutes: +150 °C (+302 °F)
PMP51: –40 to +130 °C (–40 to +266 °F)
For a maximum of 60 minutes: +150 °C (+302 °F)
PMP55: –70 to +400 °C (–94 to +752 °F)
(depending on the filling oil)
Ambient temperature range
Without LCD: –40 to +85 °C (–40 to +185 °F)
With LCD display: –20 to +70 °C (–4 to +158 °F)
(extended temperature application range–40 to +85 °C (–40 to +185 °F) with limitations in optical
properties, such as display speed and contrast)
Separate housing:–20 to +60 °C (–4 to +140 °F)
PMP55: Diaphragm seal systems depending on the version
Reference accuracy
PMC51: up to ±0.10% of the set span
PLATINUM version: up to ±0.075% of the set span
PMP51: up to ±0.10% of the set span
PLATINUM version: up to ±0.075% of the set span
PMP55: up to ±0.10% of the set span
Supply voltage
11.5 to 45 V DC (versions with plug-in connection 35 V DC)
For intrinsically safe device versions: 11.5 to 30 V DC
Output
4 to 20 mA, 4 to 20 mA with superimposed HART protocol, PROFIBUS PA or FOUNDATION Fieldbus
Options
Inspection certificate 2.2 or 3.1 or other certificates
3A approval and EHEDG approval
Specific firmware versions
Initial device settings
Separate housing
Broad range of accessories
NACE-compliant materials
Specialties
PMC51:
Metal-free measurement with PVDF connection
Special cleaning of the transmitter to remove paint-wetting impairment substances, for use in
paint shops
PMP51:
Process connections with minimum oil volume
Gas-tight, elastomer-free
PMP55:
Wide range of diaphragm seals
For extreme medium temperatures
Process connections with minimum oil volume
Fully welded versions