18 SuperUser mode
18.11 Display of the last valid measured value FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
184
18.11 Display of the last valid measured value
If the signal is not sufficient for a measurement, UNDEF is normally displayed. Instead of
UNDEF, it is also possible to display the last valid measured value.
• Select Special Funct.\SYSTEM settings\Measuring\Miscellaneous.
• Press ENTER until the menu item Keep display val is displayed.
• Select on to activate the display of the last valid measured value, off to deactivate it.
• Press ENTER.
18.12 Displays during the measurement
Besides the normal information, see section 9.4, the following parameters can be dis-
played during the measurement in the SuperUser mode:
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Keep display val
display explanation
t= transit time of the measuring signal
c= sound speed
REYNOLD= Reynolds number
VARI A= standard deviation of the signal amplitude
VARI T= standard deviation of the transit time of the measuring signal
dt-norm= transit time difference standardized to the transducer frequency
density of the fluid