18 SuperUser mode
18.11 Display of the last valid measured value FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
184
18.11 Display of the last valid measured value
If the signal is not sufficient for a measurement, UNDEF is normally displayed. Instead of
UNDEF, it is also possible to display the last valid measured value.
Select Special Funct.\SYSTEM settings\Measuring\Miscellaneous.
Press ENTER until the menu item Keep display val is displayed.
Select on to activate the display of the last valid measured value, off to deactivate it.
Press ENTER.
18.12 Displays during the measurement
Besides the normal information, see section 9.4, the following parameters can be dis-
played during the measurement in the SuperUser mode:
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Keep display val
display explanation
t= transit time of the measuring signal
c= sound speed
REYNOLD= Reynolds number
VARI A= standard deviation of the signal amplitude
VARI T= standard deviation of the transit time of the measuring signal
dt-norm= transit time difference standardized to the transducer frequency
density of the fluid

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F60 Series and is the answer not in the manual?

Flexim FLUXUS F60 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F60 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals