18 SuperUser mode
FLUXUS F60* 18.6 Detection of long measurement failures
181
UMFLUXUS_F60xV5-0EN, 2017-10-16
18.6 Detection of long measurement failures
If there are no valid measured values during a long time interval, new increments of the
totalizers will be ignored. The values of the totalizers remain unchanged.
In the SuperUser mode, it is possible to set the time interval.
Select Special Funct.\SYSTEM settings\Measuring\Miscellaneous.
Press ENTER until the menu item Do not total. if no meas. is displayed.
Enter the time. If zero is entered, the default value of 30 s will be used.
Press ENTER.
18.7 Number of decimal places of the totalizers
The totalizer values can be displayed with up to 11 places, e.g., 74890046.03. In the
SuperUser mode it is possible to define the number of decimal places.
Select Special Funct.\SYSTEM settings\Measuring\Miscellaneous.
Press ENTER until the menu item Total digits is displayed.
Select one of the following list items:
Automatic: dynamic adaptation
Fixed to x digit: x decimal places (range: 0...4)
Press ENTER.
Total digits = Automatic
The number of decimal places will be adjusted dynamically. Low totalizer values will ini-
tially be displayed with 3 decimal places. If the values of the totalizers are higher, the
number of decimal places will be reduced.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Do not
total. if no meas.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Total digits
max. value display
< 10
6
±0.000 ±999999.999
< 10
7
±1000000.00 ±9999999.99
< 10
8
±10000000.0 ±99999999.9
< 10
10
±1000000000 ±9999999999

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F60 Series and is the answer not in the manual?

Flexim FLUXUS F60 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F60 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals