18 SuperUser mode
18.5 Upper limit of the sound speed FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
180
Enter for each measuring channel the offset. Enter zero to use the default value of
300 m/s.
Press ENTER.
It is possible to display the valid range for the sound speed (SS=) in the lower line within
the program branch Measuring during the measurement. The second value (here:
2146 m/s) corresponds to the upper limit at the operating point.
Special Funct.\SYSTEM settings\Measuring\Miscellaneous\Bad
soundspeed\offset
Example
fixed upper value of the sound speed thresh.: 2007 m/s
offset: 600 m/s
value of the sound speed curve at the operating point: 1546 m/s
As 1546 m/s + 600 m/s = 2146 m/s is greater than the fixed upper value 2007, this val-
ue will be used as the upper limit of the sound speed when the plausibility of the signal
is evaluated.
Fig. 18.1: Display of the valid sound speed range
Notice!
The defined upper limit of the sound speed remains active after the deactivation of
the SuperUser mode.
GAIN=91dB
SS=1038/2146 m/s

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F60 Series and is the answer not in the manual?

Flexim FLUXUS F60 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F60 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals