8 Start-up
8.6 Initialization FLUXUS F60*
2017-10-16, UMFLUXUSF60xV5-0EN
78
8.6 Initialization
During an initialization (INIT) of the transmitter, the settings in the program branches
Parameter and Output Options and some of the settings in the program branch
Special Funct. are reset to the default settings of the manufacturer.
Proceed as follows to execute an initialization:
When switching on the transmitter: keep keys BRK and C pressed.
During the operation of the transmitter: press keys BRK, C and ENTER at the same
time. A reset is executed. Release only key ENTER. Keep keys BRK and C pressed.
After the initialization has been executed, the message INITIALISATION DONE is dis-
played.
After the initialization, the remaining settings of the transmitter can be reset to the default
settings and/or the stored measured values can be deleted.
Select yes to reset the remaining settings of the transmitter to default settings or no to
keep them at the current settings.
Press ENTER.
If yes is selected, the message FACTORY DEFAULT DONE will be displayed.
Select yes to delete the stored measured values or no to keep them.
Press ENTER.
This display will only be indicated if measured values are stored in the transmitter.
FACTORY DEFAULT
Delete Meas.Val.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F60 Series and is the answer not in the manual?

Flexim FLUXUS F60 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F60 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals