19 Settings
FLUXUS F60* 19.2 Measurement settings
191
UMFLUXUS_F60xV5-0EN, 2017-10-16
Select on to store and to output the values of the heat quantity totalizer and the volume
totalizer during heat flow measurement.
Press ENTER.
Select the overflow behavior of the totalizers, see section 17.1.
Press ENTER.
Select on to keep the previous totalizer values after a restart of the measurement.
Select off to reset the totalizers to zero after a restart of the measurement.
Press ENTER.
It is possible to set a time period after which the display toggles between the positive and
the negative totalizer during the measurement.
Select a time between 0 (deactivated) and 5 s.
Press ENTER.
The activation of the turbulence mode can enhance the signal quality if the flow is highly
turbulent (e.g., in the vicinity of an elbow or valve). An SNR value of min. 6 dB is required
during the measurement.
Select on to activate the turbulence mode.
Press ENTER.
Special Funct.\...\Measuring\heat+flow quant.
Special Funct.\...\Measuring\Quant. wrapping
Special Funct.\...\Measuring\Quantity recall
Special Funct.\...\Measuring\Toggle totalizer
Special Funct.\...\Measuring\Turbulence mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim FLUXUS F60 Series and is the answer not in the manual?

Flexim FLUXUS F60 Series Specifications

General IconGeneral
BrandFlexim
ModelFLUXUS F60 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals