Multifunction Electrical Tester Calibrator
Safety Information
3
Do not touch voltages >30 V ac rms, 42 V ac peak, or
60 V dc.
Do not use the Product if it operates incorrectly.
Do not apply more than the rated voltage, between the
terminals or between each terminal and earth ground.
Use only high-quality, shrouded test leads and adapters
with the appropriate voltage rating between the Product and
instruments being calibrated.
Do not connect to live output terminals. The Product can
supply voltages that can cause death.
Keep hands away from all Product terminals during
operation. Lethal voltages can be present on terminals.
Do not connect a hipot or insulation resistance voltage
source to the Product terminals that can source more than
100 mA.
Limit operation to the specified measurement category,
voltage, or amperage ratings.
Use the correct terminals, function, and range for
measurements.
Use only cables with correct voltage ratings.
Remove all probes, test leads, and accessories that are not
necessary for the measurement.
Do not touch exposed metal on banana plugs, they can have
voltages that could cause death.
When appropriate cable adapters are not available and
single test leads are used for calibration, hazardous
voltages can exist on the test leads. Test leads and
connectors should not be touched while the Product is in
the Operate mode.
Whenever the nature of the operation permits, keep one
hand away from equipment to reduce the hazard of current
flowing through vital organs of the body.
Make sure the Product is in Standby while making
connections to the OUTPUT HI and LO or the ZL, ZGND,
RCD terminals. Lethal voltages can be present on these
connectors during Operate mode.