EasyManua.ls Logo

Fluke Calibration 5322A - Calibrate Continuity Testers

Fluke Calibration 5322A
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multifunction Electrical Tester Calibrator
DUT Calibration Examples
165
Calibrate Continuity Testers
XW Warning
To avoid electric shock, use only high-quality, shrouded test
leads and adapters with the appropriate voltage rating between
the Product and instruments being calibrated.
Continuity is a low-ohms test function typically found on many electrical testers,
including insulation testers and installation testers. To perform a 2-wire
resistance calibration:
1. Push .
2. Using Figure 30 for a reference, connect the DUT to the Product’s LOΩ HI
and LO terminals.
3. Push the MODE softkey. Then, using the cursor keys or rotary knob, highlight
Resistance 2-Wire and select it by pushing the Select softkey or pushing in
the rotary knob.
4. In the OUTPUT area of the display, verify 2-Wire is displayed. If not, push the
MODE softkey and follow the instructions in step 3 above to select
Resistance 2-Wire.
5. Adjust the output value to the desired resistance.
6. Push .
The resistance is applied to the output terminals. Compare the reading on the
DUT with the standard value on the Product’s display.
Fluke 1653
L (L1/Red)
Fluke 5322A
PE
(L2/Green)
iep030.eps
Figure 30. 2-Wire Resistance Calibration DUT Connections

Table of Contents

Other manuals for Fluke Calibration 5322A

Related product manuals