Multifunction Electrical Tester Calibrator
Calibrate Load Test for Appliance Testers
197
Calibrate Load Test for Appliance Testers
Use the Product Meter function to calibrate an appliance tester’s voltage and
current meters. To perform a load test calibration:
1. Push .
2. Connect the DUT to the Product’s Meter A, COM and V terminals as in
Figure 55 and Figure 56. Also, make a protected earth (ground) connection
between the DUT and the Product’s PE terminal.
3. Set ground mode to OFF through the Setup softkey.
4. On the DUT, push Start.
The DUT connects power line voltage to its test socket. The DUT and
Product measure power line consumption and load current.
5. Compare the DUT reading with the Product reading.
I
PE
< 3
.5
mA
(0.
75mA P
O
RTAB
LE
AP
PLI
A
N
C
E
)
I
NS
ULAT
ION
I SU
B
BO
ND
ZE
R
O
IEC
LE
A
D
EAR
T
H
BOND
LO
AD
/ L
EA
K
I
T
O
UC
H
I
T
OU
CH
<
0.
25 mA
R
ISO
> 1.0
M C
lass I R
ISO
>
2.0
M Cl
a
ss II
RPE
<
0.
1
R
ISO
>
1.0
M C
las
s I
R
P
E
20
0
mA
,
R
PE
25A
R
PE
< 0.1 +
LEAD
PROB
E O
N
L
Y
W
I
TH
CLA
S
S
I
I
AP
PL
IAN
C
E
6500-2
APPLIANCE TESTER
External Load
Line Terminal
Protection Earth
Terminal
Neutral Terminal
Fluke 5322A
Fluke 6500
iep058.eps
Figure 55. Voltage and Current Meter Calibration with a Cable Adapter