Fluke Calibration 5322A - IEEE 488.2 Common Commands

Fluke Calibration 5322A
216 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Multifunction Electrical Tester Calibrator
Supported SCPI Commands
149
IEEE 488.2 Common Commands
The commands in this section are common to the IEEE 488.2 standard. A
summary of the Product IEEE Command Commands are listed in Table 17.
Instrument Identification
*IDN?
This command returns the identification of the manufacturer, model, serial
number and firmware revision.
When 5320A remote emulation is set to Off, the *IDN? response is
"FLUKE,5322A,serial number,firmware revision", for example
"FLUKE,5322A,650001217,0.045"
When 5320A remote emulation is set to On, the *IDN? response is
"FLUKE,5320A,serial number,firmware revisions", where firmware revisions are
comprised of four parts separated with a "+".
An example is "FLUKE,5320A,650001217,3.45+2.44+1.34+1.01".
Operation complete
*OPC
This command sets the OPC bit in the ESR (Event Status Register) when all
pending operations are complete.
Operation complete?
*OPC?
This command returns 1 to the output queue after all pending operations are
complete.
Option Identification
*OPT?
This command returns the Product hardware configuration.
The reply is formatted as follows: o1,o2,o3,x,x,x
Where:
The value of o1 references the calibrator module
The value of o2 references the compensator module
The value of o3 references the 5kV resistance decade
Note
The last three positions in the return string are reserved for future
use.
A value of "0" indicates the option is not present. A value of "1" indicates the
option is present.
Example: 1,1,1,0,0,0 means the calibrator, compensator, and 5kV resistance
decade are present.