Operators Manual
148
STAT:OPER:COND?
Description: Returns the content of the Operational Condition register. It is a
decimal value which corresponds to the binary-weighted sum of all
bits in the register. The register is not cleared after this query. The
response to the query therefore represents an instantaneous
Snapshot of the register state, at the time that the query was
accepted.
Query: STAT:OPER:COND?
STAT:QUES:EVEN?
Returns the content of Questionable Data Event register. It is a decimal value
which corresponds to the binary-weighted sum of all bits set in the register.
Register is cleared after this query.
Query: STAT:QUES:EVEN?
STAT:QUES:ENAB? <DNPD>
This command enables bits in the Questionable Data Enable register. Selected
bits are summarized at bit 3 (QSS) of the IEEE488.2 Status Byte register.
Parameters: <DNPD> Parameter represents desired value
of Questionable Data Enable
register bits as a decimal number.
Product returns the value of the
register as a decimal number.
Example: 64 is returned as 64.
Query: STAT:QUES:ENAB 2 Set bit 1 in the Questionable Data
Enable register (other bits are 0).
STAT:QUES:COND?
Returns the content of Questionable Condition register. It is a decimal value
which corresponds to the binary-weighted sum of all bits in the register. The
register is not cleared after this query. The response to the query therefore
represents an instantaneous Snapshot of the register state, at the time that the
query was accepted.
Query: STAT:QUES:COND?
STAT:PRES
This command clears all bits in the Operation Data Enable register and in the
Questionable Data Enable register.
Example: STAT:PRES