Multifunction Electrical Tester Calibrator
Supported SCPI Commands
105
RCDP – RCD PAT mode
LIN – Line Impedance mode
LOOP – Loop Impedance mode
VOLT – Voltage Calibrator mode
MET – Multimeter mode
HIPL – Hipot Leakage current
HIPT – Hipot Timer
FLI – Flash Class I
FLII –Flash Class II
[SOUR]:SAF:GBR[:LEV] (?) [<DNPD>]
Description: This command sets the resistance value. If not already selected,
this command switches the Product to the Ground Bond Resistance
function.
Parameters: <DNPD> Parameter represents the value of the resistance
expressed in Ohms. The Product selects the
nearest resistance value.
Example: SAF:GBR 0.1 Ground Bond Resistance with 100 mΩ.
Query: SAF:GBR? Product returns the set value of resistance using
exponential format. Example: 50.54mΩ is
returned as 50.54e-03.
[SOUR]:SAF:GBR:RPOS (?) <DNPD>
Description: This command selects one of the 16 available resistances. Index of
resistance is from 0 (25 mΩ) to 15 (2 kΩ). If not already selected,
this command switches the Product to the Ground Bond Resistance
function.
Parameters: <DNPD> Parameter represents index of required
resistance. Acceptable range is 0 … 15.
Example: SAF:GBR:RPOS 2 Ground Bond Resistance index 2 (100mΩ).
Query: SAF:GBR:RPOS? Product returns the index of selected
resistance. Example: 4 is returned as 4.
[SOUR]:SAF:GBR:SENS(?) <CPD> { ON | OFF }
Description: This command selects 4-wire Ground Bond Resistance mode
(remote sense). If not already selected, this command switches
the Product to the Ground Bond Resistance function.
Parameters: <CPD> ON selects 4-wire mode
OFF selects 2-wire mode
Example: SAF:GBR:SENS ON 4-wire Ground Bond Resistance mode .
Query: SAF:GBR:SENS? Product returns ON in 4-wire mode and
OFF in 2-wire