5322A
Operators Manual
132
[SOUR]:SAF:RCDP:POL?
Description: Returns the type of test signal generated by DUT { POS | NEG |
SYMP | SYMN | DCP | DCN | NO }. If not already selected, this
command switches the Product into the RCD PAT function mode.
Types are:
Test current is pulse (dc) with positive polarity (POS)
Test current is pulse (dc) with negative polarity (NEG)
Test current is symmetrical (ac) with positive phase (SYMP)
Test current is symmetrical (ac) with negative phase (SYMN)
Test current is dc with positive polarity (DCP)
Test current is dc with negative polarity (DCN)
Answer NO is sent in case that no valid test signal is generated.
Query: SAF:RCDP:POL?
[SOUR]:SAF:RCDP:REC <CPD> { ON | OFF }
Description: This command automatically re-connects the output terminals and
places the Product into Operate after tripping. The re-connection
occurs in a fixed time interval of approximately 2.5 seconds.
Parameters: <CPD> ON enables Automatic reconnection
OFF disables Automatic reconnection
Query: SAF:RCDP:REC?
[SOUR]:SAF:LIN [:LEV](?) [<DNPD>]
Description: This command switches the Product into the Line Impedance
function mode and sets the resistance value.
Parameters: <DNPD> Parameter represents the value of the
resistance expressed in Ohms. The Product
selects the nearest resistance value.
Example: SAF:LIN 0.1 Line impedance of 100 mΩ.
Query: SAF:LIN? The Product returns the set value of
resistance using exponential format.
Example: 900 Ω is returned as 900.0e+00.