Remote Operation
Supported SCPI Commands 5
5-13
Table 5-4. SOURce Command Summary (cont.)
Command Description
[:CURRent]:LEVel? Returns the I level value.
[:VOLTage]:LINE? Returns the power supply voltage.
[:VOLTage]:TOUCh? Returns the measured touch voltage.
:POLarity? Returns type of test signal generated by UUT.
:RCDC Path to RCD Trip Current mode.
:CURRent? Returns the measured trip current.
:RPOSition? Returns the index value of the selected resistance.
[:CURRent]:NOMinal <DNPD> Sets the nominal current value for RCD Trip Current
mode.
[:CURRent]:NOMinal? Returns the nominal trip current value.
[:VOLTage]:LINE? Returns the power supply voltage.
[:VOLTage]:TOUCh? Returns the measured touch voltage.
:POLarity? Returns type of test signal generated by UUT.
:LINeimpedance Path to Line Impedance mode.
[:LEVel] <DNPD> Sets the resitance value for the Line Impedance mode.
[:LEVel]? Returns the resistance value set for the Line Impedance
mode.
:CORRection {OFF|MAN|SCAN|COMP} Sets the mode of residual impedance correction.
:CORRection? Returns the residual impedance correction mode.
:POLarity? Returns type of test signal generated by the UUT.
:CURRent? Returns measured current flowing through resistance.
:CLEar Clears the maximum value of peak current.
[:CURRent]:PFC? Returns Prospective Fault Current value.
:MANual <DNPD> Sets the Manual residual line impedance value.
:MANual? Returns the set Manual residual line impedance value.
:SCAN? Returns Scanned line residual impedance value.
:LOOPimpedance Path to Loop impedance mode.
[:LEVel] [<DNPD>] Sets resistance value for the Loop impedance mode.
[:LEVel]? Returns the set resistance value for Loop impedance
mode.
:CORRection {OFF|MAN|SCAN|COMP} Sets mode of residual impedance correction.
:CORRection? Returns the set mode of the residual impedance
correction.
:POLarity? Returns type of test signal generated by the UUT.

Table of Contents

Other manuals for Fluke 5320A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5320A and is the answer not in the manual?

Fluke 5320A Specifications

General IconGeneral
BrandFluke
Model5320A
CategoryTest Equipment
LanguageEnglish

Related product manuals