Two Jumper Reference Method
This method is used when the link includes:
• the loss of one connection at the near or far end cross connect
• the loss of the cable plant
TIA/EIA -455-171and IEEE 10 Base-X specify the two jumper method.
Note
This method is called ‘Method A’ in TIA-526-14A and ‘Method A.2’
in TIA-526-7.
1. Connect a launch cable to the fiber adapter that is attached to
OMNISCANNER. The black fiber attaches to the TX (transmitter) port and
the white fiber to the RX (receiver) port.
2. Connect a launch cable to the fiber adapter that is attached to
OMNIREMOTE. The black fiber attaches to the TX (transmitter) port and
the white fiber to the RX (receiver) port.
3. Use a connector adapter to mate the two black fiber connectors.
4. Use a connector adapter to mate the two white fiber connectors.
5. Press (Setup) to open the SETUP screen and then select Set
Reference to set the reference.
6. Disconnect the connectors from the connector adapters without
disturbing the fiber connections to the OMNIFibers.
7. Set the connector adapters aside.
2 m Launch Cable
2 m Launch Cable
Coupler