OMNISCANNER User Guide
3 - 4
OMNISCANNER will store reference data for up to 5 OMNIFIBER pairs.
Note
Once the reference value has been set, ensure that both launch cables
remain attached to the fiber adapters’ TX (transmit) ports so that the
connections are not disturbed.
6. Once the reference value has been set, OMNISCANNER will display the
FIBER INFO screen.
7. Press ESC to return to the SETUP screen.
Project Selection
A project name must be selected for test result management. Autotests
and Cables were previously assigned to a specific project through the
use of the OMNISCANNER Configuration software. (See Chapter 9 -
Uploading and Printing for further information about OMNISCANNER
Configuration.)
The Project feature is very useful for grouping
test results and eases uploading and sorting of
Autotests stored in the scanner.
1. Use the é or ê to highlight Project: and
press ENTER.
2. Select the appropriate Project from the pop-
up list and press ENTER.
3. Use the é or ê to highlight Autotest: and
press ENTER.
Note
Different Autotests have different reference
methods already assigned to them.
4. Select the appropriate Autotest from the
pop-up list and press ENTER.
5. Use the é or ê to highlight Cable: and
press ENTER.
6. Select the cable type to be tested from the pop-up list and press
ENTER.