OMNISCANNER User Guide
Three Jumper Reference Method
This method is used when the link includes:
• the loss of the cable plant
Note
This method is called ‘Method C’ in TIA-526-14A and ‘Method A.3’
in TIA-526-7.
1. Connect the first launch cable to the fiber adapter that is attached to
OMNISCANNER. The black fiber attaches to the TX (transmitter) port and
the white fiber to the RX (receiver) port.
2. Connect the second launch cable to the fiber adapter that is attached
to OMNIREMOTE. The black fiber attaches to the TX (transmitter) port
and the white fiber to the RX (receiver) port.
3. Use two connector adapters to mate the black TX connector to the
white RX connector through a third jumper cable.
4. Use two connector adapters to mate the other black TX connector to
the white RX connector through a third jumper cable.
5. Press (Setup) to open the SETUP screen and then select Set
Reference to set the reference.
6. Remove the third jumper cable, located in the center, and the
connector adapters without disturbing the OMNIFIBER connections.
2 m Launch Cable
Couplers
2 m Launch Cable 2 m Launch Cable