A - 7
OMNISCANNER User Guide
Length (OMNIFIBER MM, multimode)
· Maximum 2000m (6600ft)
· Resolution: 1m (1ft)
Length (OMNIFIBER SM, single-mode)
· Maximum 16,000m (52,500ft)
· Resolution: 1m (1ft) from 0 to 2000m, 2m (6ft) from 2,000 to 16,000m
Delay (OMNIFIBER MM, multimode)
· 0 to 10000nS
· 1nS resolution 0-10,000nS
Delay (OMNIFIBER SM, single-mode)
· 0 to 80,000nS
· 1nS resolution 0-10,000nS; 8nS resolution 10,000 - 80,000nS
High Voltage Input Protection
· The test interface withstands input hazard conditions that arise from
normal telephone interfaces (48 VDC at less than 80 mA) or 24 VAC
power used to power many telephone keysets. The unit detects when
a voltage of greater than ±15 volts is applied on the inputs and
sounds a warning to alert the user to remove the hazard condition
immediately to prevent possible damage to the input circuits.
Note
Measurements cannot be performed when hazard conditions are
present.