Chapter 4. Initial Setup and Programming
56 AquaTrans™ AT600 User’s Manual
4.6.2.3 Setting the Frequency
Setting Measurement Type
Setting the Base Value, Full Value, and Frequency
Frequency sends out a continuous square wave, with a frequency
proportional to the measured flow rate. Select
Frequency and press [] to
open the following screen.
Select
Measurement and press [] to open a screen similar to the
following.
In the
Measurement menu, select the type of analog output to be used and
press
[] to return to the previous screen.
The
Base Value is the measurement that corresponds to a 0 Hz pulse. The
Full Value is the measurement value that corresponds to a Full Frequency
pulse. The
Full Frequency is the maximum frequency used for the output
pulse, and it indicates the maximum flow rate measurement. Use the []
or [] arrow keys to select the appropriate option and press
[]. A screen
similar to the following opens.
Note: Use the same steps to program the
Base Value, Full Value and Full
Frequency
.
To change the existing number, press
[] and a screen similar to the
following opens.
Note: The units shown are the units selected in the
Units Setting menu.
Digital Output B
Off
Pulse
Frequency
Alarm
Frequency
Measurement
Base Value
Full Value
Full Frequency
Velocity
Act Volumetric
Std Volumetric
Mass
Measurement
Base Value
Full Value
Base Value
Base Value
0.0000 m/s