AquaTrans™ AT600 User’s Manual 57
Chapter 4. Initial Setup and Programming
Setting the Base Value, Full Value, and Frequency (cont.)
Setting the Frequency Error Handling
Use the [] or [] key to select a specific digit, then use the [] or [] key
to change that digit as required. After all digits are correct, press
[] to
save. Then, press [
] to return to the previous screen.
To change the
Error Handling status, select Error Handling in the menu and
press
[]. The following screen opens.
To change the current
Error Handling status, select the desired option, and
press
[]. The screen returns to the previous display.
There are four options for response to an error situation:
• Hold: hold the last good value
• Low: display 0 Hz
• High: display the Full Frequency
• Other: If selected, a screen similar to the following opens
Enter the frequency value you want to display during an error situation.
(e.g., if
Full Frequency = 1 kHz, you may want to set the Error Handling Value
to 2 kHz.) Press [] again, and a screen similar to the following opens.
Use the [] or [] key to select a specific digit, then use the [] or [] key
to change that digit as required. After all digits are correct, press
[] to
save. Then, press [
] to return to the previous screen.
Base Value
UNIT: m/s
0.000
]MOVE
]UNDO [ ]SAVE
]
]
]MODF
[
Frequency
Full Value
Full Frequency
Test Frequency
Error Handling
Error Handling
Low
High
Hold
Error Handling Value
Value
0 Hz