Revision C CASE 1-7
2060290-201
Safety Information
Warning
Shock Hazard — Observe the following warnings. Failure
to do so endangers the lives of the patient, the user, and
other persons present.
Only specially trained technical staff is authorized to
open the connector access panel.
Before putting the device into operation, the operator
is required to ascertain that it is in correct working
order and operating condition. The cables, in
particular, must be checked for damage. Damaged
cables and connectors must be replaced immediately.
When disconnecting the device from the power line,
remove the plug from the wall outlet, before
disconnecting the cable from the device. Otherwise
there is a risk of coming in contact with line voltage
by inadvertently introducing metal parts in the
socket of the power cord.
Do not use extension cables with multiple portable
socket outlet (MPSO).
In the patient vicinity, the printer installed on the
system cart must meet the requirements of IEC
60601, or it must be a GE Healthcare modified
printer with a connected protective earth conductor
(potential equalization). Observe the installation
instructions supplied with the printer.
All devices of a system must be connected to the same
power supply circuit. Devices that are not connected
to the same circuit must be electrically isolated for
operation (electrically isolated RS 232 interface); this
requirement does not apply in the USA.
The device is not intended for intracardiac use.
All bicycle ergometers and treadmills connected to
the system must meet the requirements of IEC
60601-1. CASE and the connected bicycle ergometer
or treadmill must be connected to the same electric
circuit via separate wall outlets; this requirement
does not apply in the USA.