1-8 CASE Revision C
2060290-201
Safety Information
Warning
Shock Hazard — Observe the following warnings. Failure
to do so endangers the lives of the patient, the user, and
other persons present.
All modems connected to the system must meet the
requirements of IEC 60950 or UL1950. The specific
regulations valid in your country must also be
observed.
The modem must be installed within the medical
location, but not in the patient vicinity.
Disconnect the patient from the system while
sending data via the modem.
Devices may only be connected with each other or to
parts of systems when it has been made certain that
there is no danger to the patient, the operator, or the
environment as a result. In those instances where
there is any element of doubt concerning the safety of
connected devices, the user must contact the
manufacturers concerned or other informed experts
as to whether there is any possible danger to the
patient, the operator, or the environment as a result
of the proposed combination of devices. Standards
IEC 60601-1-1/EN 60601-1-1 must be complied with
in all cases.
For connection of the system cart to a network, a
transceiver according to IEC 60601-1 or IEC 60601-1-
1 must be used.
Liquids must not be allowed to enter the device. If
liquids have entered the device, take it out of service
and have it checked by a service technician before
using it again.
Refer servicing to authorized service personnel. Do
not open the device. There are no user-replaceable
components inside the device.
Do not insert objects of any kind into the slots of the
housing. They may touch live components and you
might suffer an electric shock, cause fire, or damage
the device.