Chapter 5: Repair Procedures
Probe Panel Housing Repairs
Service Manual 5-19
5.5.6 Probe Panel Upper Housing Replacement
1. Follow procedures 5.5.1 through 5.5.4 as appropriate to remove all parts from the Probe Panel Upper
Housing.
FIGURE 5-10. Probe Panel Assembly (with SET 1 SpO
2
Jack)