Appendix D: Troubleshooting
Service Mode (Application Software 1.0.9 and Earlier)
D-2 Service Manual
D.1.1.1 Unit Information Displays
• Unit type
•Software
• Processor and option versions
• PC Board ID’s
• Battery voltage status
• Power fail minutes
• Language
D.1.2 Operator Settings
Operator Settings enable users to customize the behavior of alarms, user functions, and displayed units by
setting default values for the parameters and options listed in the following table.
Parameter/Option Settings Default
Patient temp alarm 0.5C / 1.0C 1.0C
CPAD (check patient alarm disabled) 5% - 50% 25%
Man temp alarm Off / 0.5C / 1.0C Off
Touchless silence Enable / Disable Enable
Bed elevate (for Bedded Warmers with
Elevating Base)
Enable / Disable /
Disengaged
Enable
Bed elevate (for Bedded Warmers with
Fixed-Adjustable Base, Freestanding
Warmers, and Wall Mount Warmers)
Enable / Disable /
Disengaged
Disengaged
Temp Units C / F C
Scale Units grams / lbs-oz grams
Scale Resolution 10/5 grams 10 grams
Time Format 12/24 24
Low pulse limit 0 - 110 100
High pulse limit 111 - 240 200
Low sat limit 0 - 99 85
High sat limit 86 - 105 100
Masimo SpO
2
averaging 2-4 / 8 / 16 8
Masimo SpO
2
sensitivity Normal / Maximum Normal
Nellcor SpO
2
Response Mode Normal / Fast Normal
Nellcor SpO
2
Display Pleth / Blip Pleth
Pulse tone volume, Panda unit 0/ 1 / 2 / 3 / 4 1
Pulse tone volume, Giraffe unit 0/ 1 / 2 / 3 / 4 0
Alarm volume 1 / 2 / 3 / 4 Default is 4. 3
Language Many English