Chapter 4: Troubleshooting
PC Command Line Interface
Service Manual 4-15
4.2.3 PC Command Line Interface Sub-Menus
4.2.3.1 Version
Version has no help submenu.
4.2.3.2 getbuttons
command: getbuttons
arguments: none
function:
Returns the current state (released or pushed) of the touch panel push buttons. The buttons are
numbered from 1 to 11, starting with the alarm silence button and progressing counter-clockwise. For
example, button1 = alarm silence, button5 = lower left button, button6 = ? button, and button11 = lower
right button.
usage: getbuttons
4.2.3.3 update
Note: Update instructions, command variables, and materials will be provided with the update package.
command: update
arguments:
‘proc’ = processor to update
audio - audio processor
heat - heat control processor
‘filesz’ = size of the file to upload
‘crc’ = crc of the file to upload
function:
Updates the specified processor with the file provided via xmodem upload. After issuing this command,
the user will be prompted to send the update file via xmodem transfer. Once the file transfer is
complete, the crc is checked to ensure data integrity; then the processor update (re-programming)
process will begin. The touch panel buttons will be disabled while caring out the functions of this
command.
usage: update proc filesz crc
Caution: Do not run the update command unless specifically instructed to do so.
Caution: A failed exit status of this command after commencement of processor re-
programming may leave the system in an unknown state. If this happens, the
command should be run again until successful, or the warmer should be placed out
of service and technical support contacted.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C

Related product manuals