Chapter 1: Functional Description
RS-232 Serial Data
Service Manual 1-27
1.4 RS-232 Serial Data
Note: In the event of a power failure, all serial communication will cease until power is restored.
1.4.1 RS-232 Connector
The Nurse Call and the serial data output share the same female, nine pin, d-type connector.
Pin 2: Receive Data (incubator input)
Pin 3: Transmit Data (incubator output)
Pin 5: Gnd (Signal Ground)
1.4.1.1 Cable Requirements
The user interface cable must have capacitance less than 1500 pF. It should be a shielded cable such as
Belden 9611 with AMP shielding kit 748046-1 and ferrule 747579-8.
1.4.2 Data Stream
The data stream from the Panda and Giraffe products is repeated every two seconds. The RS-232
parameters are 116000 bps baud rate, no parity, 8 data bits, and one stop bit. The data is in ASCII format;
the string is described below:
PWRM_2.0.0,36.8,36.4,0,P,0,0,0,00,29
Warning: The computer or RS-232 monitor’s user program must continuously check the data
link. The program should constantly verify connection to the control panel and check for
updated data.
HYB This is the product code.
GWRM = Giraffe Warmer
PWRM = Panda Warmer
PNBW = Panda Non-Bedded Warmer
2.0.0 This is the control board software revision (thermoregulation in the startup screen).
36.8 Temperature from Patient Probe #1.
OPEN Temperature from Patient Probe #2. OPEN represents no probe connected.
36.4 Baby (Servo) mode desired temperature setting.
0Heater power.
P Baby (Servo) Mode = P
Manual Mode = M

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C