Chapter 4: Troubleshooting
Service Mode
Service Manual 4-3
4.1.4 Operation Settings
This menu enables users to customize the behavior of alarms, user functions, and displayed units by setting
default values for the parameters and options listed in the following table.
* The Low pulse rate limit can be set only between 25 and the High pulse rate limit -1.
The High pulse rate limit can be set only between the Low pulse rate limit +1 and 240.
Feature Available Settings US Defaults International Defaults
Patient temp. alarm 0.5°C / 1.0°C 1.0°C 1.0°C
Check patient alarm disabled 5% - 50% 25% 25%
Man. temp. alarm Off / 0.5°C / 1.0°C Off Off
Alarm volume 0 / 1 / 2 / 3 3 3
Hands free silence Enable / Disable /
Disengaged
Enable Enable
Bed elevate (for Bedded Warmers
with Elevating Base)
Enable / Disable /
Disengaged
Enable Enable
Bed elevate (for Bedded Warmers
with Fixed-Adjustable Base,
Freestanding Warmers, and Wall
Mount Warmers)
Enable / Disable /
Disengaged
Disengaged Disengaged
Temp. units C / C Only / F C C Only
Scale units grams / lbs-oz /
grams only
grams grams only
Scale resolution 10/5 grams 10 grams 10 grams
Language Many English English
Low pulse rate limit* 25 - 110 100 100
High pulse rate limit* 111 - 240 200 200
Low saturation limit** 20 - 99 85 85
High saturation limit** 86 - 100 100 100
SpO
2
alarm suspension on startup 0 / 15 / 30 15 15
SpO
2
averaging (Masimo only) 2-4 / 8 / 16 8 8
SpO
2
sensitivity (Masimo only) Normal / Maximum Normal Normal
SpO
2
SatSeconds (Nellcor only) Off, 10, 25, 50, 100 Off Off
SpO
2
Response Mode (Nellcor only) Normal / Fast Normal Normal
SpO
2
Display (Nellcor only) Pleth / Blip Pleth Pleth
SpO
2
View Large / Normal Large Large
Pulse tone volume, Panda 0/ 1 / 2 / 3 / 4 1 1
Time Format 0/ 1 / 2 / 3 / 4 12 24
Date Format YYYY/MM/DD MM/DD/YYYY DD/MM/YYYY
MM/DD/YYYY
DD/MM/YYYY
APGAR TONES (Giraffe Only) ON/OFF OFF OFF