C-1
Appendix C: Alarm Priorities
Priority Alarm Subject
Activation
Criteria
Message
Displayed
Silence
Time
Sound:
1= High
2=Low
3 = informational
Notes/
Comments
Indication
Light
Special
Requirements
1SpO
2
lost
communica-
tions
SpO
2
board
has not
communicated
with the main
processor for
at least
2seconds
Oximetry
System
Failure
Refer
to
notes
1 When this alarm
occurs, the SpO
2
module will be
automatically
turned off and
SpO
2
options will
no longer be
available. When
silenced, the
alarm will clear.
Primary
2 Pulse rate PR < 100 (floats
with user
adjustable
limits)
Low Pulse
Rate
3 min. 1 Primary
3Oxygen
saturation
SpO
2
<85
(floats with
user adjustable
limits)
Low SpO
2
3 min. 1 Primary
4 Check baby Radiant %
power @ 100%
for >15 min
and alarm not
silenced
Check
Baby -
Heat Off
15 min 1 Primary Operational in
Baby Control
mode
5 Check baby Radiant heater
% above
“Check Baby
Alarm
Disabled”
(CPAD) limit for
>15 min and
alarm not
silenced
Check
Baby -
Heat Off
15 min 1 Primary Operational in
Baby Control
Mode
6Baby
temperature
>2.0° from
Control Temp
Baby Hot-
Check
Temp
Probe
5 min 1 With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 1.8°C
Primary Operational in
Baby Control
Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C