EasyManuals Logo

GE PANDA Service Manual

GE PANDA
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #186 background imageLoading...
Page #186 background image
Appendix C: Alarm Priorities
C-2 Service Manual
7Baby
temperature
>2.0° from
Control Temp
Baby Cold-
Check
Temp
Probe
5 min 1 With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 1.8°C
Primary Operational in
Baby Control
Mode
8 Check baby Radiant %
power @ 100%
for > 12 min
Check
Baby
12 min 2 Primary Operational in
Baby Control
Mode
9 Check baby Radiant Heater
% above
“Check Baby
Alarm
Disabled”
(CPAD) limit for
> 12 min
Check
Baby
12 min 2 Primary Operational in
Baby Control
Mode
10 Temperature
probe failure
Two
Thermistors in
a probed differ
by 0.5° C or
more.
Temp.
Probe
Failure
2 min 2 to 1
(after 30 secs.)
Primary Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
11 Disconnected
temperature
probe
No longer
getting a
reading from
the
temperature
probe
Confirm
Probe Jack
Connection
2 min 2 to 1
(after 30 secs.)
Primary Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
12 Baby
temperature
>1.0°C (>0.5°C)
from control
temp
Baby Hot -
Check
Temp
Probe
12 min 2 With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
Primary Operational in
Baby Control
Mode
13 Baby
temperature
<1.0°C (<0.5°C)
from control
temp
Baby Cold -
Check
Temp
Probe
12 min 2 With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
Primary Operational in
Baby Control
Mode
Priority Alarm Subject
Activation
Criteria
Message
Displayed
Silence
Time
Sound:
1= High
2=Low
3=informational
Notes/
Comments
Indication
Light
Special
Requirements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C

Related product manuals