Appendix C: Alarm Priorities
C-2 Service Manual
7Baby
temperature
>2.0° from
Control Temp
Baby Cold-
Check
Temp
Probe
5 min 1 With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 1.8°C
Primary Operational in
Baby Control
Mode
8 Check baby Radiant %
power @ 100%
for > 12 min
Check
Baby
12 min 2 Primary Operational in
Baby Control
Mode
9 Check baby Radiant Heater
% above
“Check Baby
Alarm
Disabled”
(CPAD) limit for
> 12 min
Check
Baby
12 min 2 Primary Operational in
Baby Control
Mode
10 Temperature
probe failure
Two
Thermistors in
a probed differ
by 0.5° C or
more.
Temp.
Probe
Failure
2 min 2 to 1
(after 30 secs.)
Primary Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
11 Disconnected
temperature
probe
No longer
getting a
reading from
the
temperature
probe
Confirm
Probe Jack
Connection
2 min 2 to 1
(after 30 secs.)
Primary Operational in
Baby Control
mode or
Manual Mode
when Manual
Temperature
alarm set
12 Baby
temperature
>1.0°C (>0.5°C)
from control
temp
Baby Hot -
Check
Temp
Probe
12 min 2 With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
Primary Operational in
Baby Control
Mode
13 Baby
temperature
<1.0°C (<0.5°C)
from control
temp
Baby Cold -
Check
Temp
Probe
12 min 2 With hysteresis to
avoid nuisance
alarm recurrence
after silencing.
Hysteresis set at
0.2°C - alarm
clears at 0.8°C
(0.3°C)
Primary Operational in
Baby Control
Mode
Priority Alarm Subject
Activation
Criteria
Message
Displayed
Silence
Time
Sound:
1= High
2=Low
3=informational
Notes/
Comments
Indication
Light
Special
Requirements