Appendix C: Alarm Priorities
Service Manual C-3
14 No SpO
2
probe
connected
Masimo or
Nellcor board
indicates no
board
connected
No SpO
2
Board
3 min 2 to 1
(after 10 secs)
Primary Only available
if SpO
2
is
running
15 Defective
SpO
2
probe
Masimo or
Nellcor board
indicates
defective probe
Check
SpO
2
3 min 2 to 1
(after 10 secs)
Primary Only available
if SpO
2
is
running
16 SpO
2
interference
detected
Masimo or
Nellcor board
indicates
interference
detected
Check
SpO
2
3 min 2 to 1
(after 10 secs)
Primary Only available
if SpO
2
is
running
17 SpO
2
probe
off baby
Probe off baby
is indicated
from the
Masimo or
Nellcor board
SpO
2
Probe
Off Baby
3 min 2 to 1
(after 10 secs)
Primary Only available
if SpO
2
is
running
18 Too much
ambient light
Masimo or
Nellcor board
indicates too
much ambient
light
Check
SpO
2
3 min 2 to 1
(after 10 secs)
Primary Only available
if SpO
2
is
running
19 Unrecognized
SpO
2
probe
Unrecognized
probe
indicated from
Masimo or
Nellcor board
Unrecog-
nized SpO
2
Probe
3 min 2 to 1
(after 10 secs)
Primary Only available
if SpO
2
is
running
20 Pulse rate PR >200 (floats
with user
adjustable
limits)
High Pulse
Rate
3 min 2 to 1
(after 10 secs)
Primary Only available
if SpO
2
is
running
21 Oxygen
saturation
SpO
2
>100
(floats with
user adjustable
limits
High SpO
2
3 min 1 Primary
22 Callback timer
complete
Current
callback timer
has reached
target time
Callback
Timer
Complete
Refer
to
notes
2 Silencing this
alarm will cause
the timer alarm
to shut off
Primary
Priority Alarm Subject
Activation
Criteria
Message
Displayed
Silence
Time
Sound:
1= High
2=Low
3 = informational
Notes/
Comments
Indication
Light
Special
Requirements

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C