Chapter 4: Troubleshooting
Service Mode
4-10 Service Manual
4.1.9 Event Logs
This menus displays system and user events like Power up, Key Presses, System Failures, etc. Information
displayed includes Date and Time, Ambient (Control Board) temperature, Event Name and Raw
(hexadecimal) data.
More detail is displayed below the list for a selected event. The user may scroll to any event in the list by
using the UP/DOWN arrow keys.
The Options Sub-menu provides a means to filter on one or more event types. The user may also filter by
specified dates and times and sort in ascending or descending order.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C