Chapter 4: Troubleshooting
Service Mode
4-4 Service Manual
** The Low saturation limit can be set only between 20 and the High saturation limit –1.
The High saturation limit can be set only between the Low saturation limit +1 and 100.
The Operation Settings menu screen appears as follows:
4.1.5 Scale Calibration
This menu displays the Scale Recalibration Count (SR) and Corrected value (SC). It also displays Scale
software revision, model, resolution code and POST (power on self test) results.
Note: For versions of Application Software 1.0.9 and earlier SR represents Scale Raw Value.
Menu selections:
Manual Cal (refer to “3 Calibration” for details).
Check Weight. Used to verify the correct weight is reported once calibration is complete.