B-1
Appendix B: Electromagnetic Compatibility
B.1 Electromagnetic Compatibility (EMC) Guidance
Safety Standards: IEC 60601-1 and IEC 60601-2-21,
EMC Standards: IEC 60601-1-2 3rd ed.
B.2 Manufacturer’s Guidance and Declaration Regarding Electronic Emissions
The Giraffe/Panda Warmer is intended for use in the electronic environment specified below. The user of
the Giraffe/Panda Warmer should ensure that it is used in such an environment.
Warning: Medical Electrical Equipment needs special precautions regarding EMC
and needs to be installed and put into service according to the EMC information
provided below.
Warning: Portable and mobile RF communication equipment can affect Medical
Electrical Equipment. Caution should be use when operating such devices around
Medical Electrical Equipment.
Warning: This equipment/system is intended for use by healthcare professionals
only. This equipment system may cause radio interference or may disrupt the
operation of nearby equipment. It may be necessary to take mitigation measures,
such as re-orienting or relocating the equipment/system or shielding the location.
Electromagnetic Emissions
Emissions Test Compliance Electromagnetic Environment Guidance
RF emissions CISPR 11 Group 1 The Giraffe/Panda Warmer uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions CISPR 11 Class A The Giraffe/Panda Warmer is suitable for use in all
establishments, other than domestic and those
directly connected to the public low voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies