A-1
Appendix A: Specifications
A.1 Power Requirements and Accessory Outlets
A.2 Operating Environment
A.3 User Control Settings
Power Requirements Accessory Outlets
5.25 A @ 100V ~, 50/60 Hz 2 A @ 100V ~, 50/60 Hz
4.57 A @ 115V ~, 50/60 Hz 2 A @ 115V ~, 50/60 Hz
2.39 A @ 220V ~, 50/60 Hz 1A @ 220V ~, 50/60 Hz
2.28 A @ 230V ~, 50/60 Hz 1 A @ 230V ~, 50/60 Hz
2.19 A @ 240V ~, 50/60 Hz 1A @ 240V ~, 50/60 Hz
Temperature 18 to 30°C
Humidity 5 to 75% Non-condensing relative humidity
Pressure 70-106 kPa
Air Velocity up to 0.3 m/s
Storage Conditions
Temperature -25 to 38°C
Humidity 0 to 85% Non-condensing relative humidity
Pressure 50 to 106 kPa
Electrical
Operation Continuous
Device Type Class 1
Applied Part Type BF
Patient control temperature 34-37.5°C in 0.1° increments
Radiant heat power 0-100% in 5% increments

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C

Related product manuals