Chapter 1: Functional Description
System Functions
Service Manual 1-15
1.3.6 External Communication
Serial communication is provided though the ThermaLink port. (Refer to Figure 1-9.) A null modem cable is
required for connection to a PC serial port. Ethernet communications are not used at this time.
A string of ASCII characters will be transmitted over the RS-232 port every 15 seconds ±5 seconds.
The string will be formatted as follows:
Unit Type with version, baby temperature, set temperature, current heater setting, mode, last recoded
weight, saturation, pulse rate, list of alarms, 00(used to note the end of the alarm list), checksum (in
hexadecimal format).
Example 1: Manual Mode with no alarms.
GWRM_1.0.7,36.3,36.5,35,M.1340,0,0,00,0d
Example 2: Baby mode with two alarms
PWRM_1.0.7,36.3,37.1,40,P,1340,0,0,16,22,00,ea
Note: Ethernet communication is not used at this time.
FIGURE 1-9. Thermalink Communication

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C

Related product manuals