Chapter 1: Functional Description
RS-232 Serial Data
1-28 Service Manual
1.4.3 Nurse Call
1.4.3.1 Contact Ratings
Maximum resistive load: 4 VA
Maximum DC switching voltage: 100 VDC
Maximum switching current: 0.25 A
Maximum carrying current: 0.50 A
1.4.3.2 Connector
The Nurse Call contacts and the serial data output share the same female, nine pin, d-type connector.
1.4.3.3 Nurse Call Signals
Pin 6: Closed contact under normal conditions, that is, power on, no alarm (recommended
configuration).
Pin 1: Common contact.
Pin 9: Open contact under normal conditions, power on, no alarm. These contacts are not
powered. They only provide closure.
Incubator Status
Nurse Call Signal
Pins 1 and 6 Pins 1 and 9
Normal Closed Open
Alarm Open Closed
Pwr switch off or pwr fails Open Closed
Nurse Call cable disconnected Open Open

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C

Related product manuals