Appendix E: Installing Wall Mount Units
Enclosure Dimensional Specification
Service Manual E-11
E.6 Enclosure Dimensional Specification
GE Healthcare does not offer an in-built bed, cabinet or other enclosure option for this product. If the facility
chooses to place the warmer in an enclosed space (e.g. cabinet, armoire), make sure that the enclosure
meets the minimum dimensional requirements as illustrated in Figure E-8 in order to have adequate air flow
to avoid warmer overheating when the enclosure is closed. During use (when the enclosure is open),
confirm that no object is blocking the radiant heat path to the bed (see Figure E-10) and that the caregiver
can have unobstructed access to the warmer control panel, probe panel, rails and the bed surface. Do not
operate the warmer unless the cabinet, armoire or other enclosure is open (see Figure E-11). GE Healthcare
will not install a warmer in an enclosure that does not provide the caregiver with unobstructed access to the
warmer control panel, probe panel, rails and the bed surface. The facility’s installation of, or operation of, the
warmer in such an enclosure is at the facility’s risk.
Warning: Do not place objects in the radiant heat path. Objects will be heated and
could block heat to the baby.
Warning: The caregiver shall have unobstructed access to the warmer control
panel, probe panel, dovetail rails and the bed surface during operation.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C

Related product manuals