Page #78 background imageLoading...
Page #78 background image
Chapter 4: Troubleshooting
PC Command Line Interface
4-16 Service Manual
4.2.3.4 logs
Refer to section 4.2.4.
4.2.3.5 pulsetone
command: pulsetone
arguments:
‘vol’ = desired speaker volume (optional)
1(min) - 4(max)
function:
Sets the volume to the specified level and generates a pulsetone.
If the argument ‘vol’ is not specified, a pulsetone will be generated at the current volume setting.
usage: pulsetone [vol]
4.2.3.6 ncas
Note: NCAS is an early acronym for Hands Free Alarm Silence.
command: ncas
arguments: none
function:
Retrieves and displays the instantaneous state of the Hands Free Alarm Silence input to the main
processor. The state is reported as ‘active’ or ‘inactive’.
usage: ncas
4.2.3.7 dips
command: dips
arguments: none
function: Reads and prints out the status of the control board DIP switches.
usage: dips
4.2.3.8 heateng
command: heateng
arguments:
‘info’ = report sw version and heater/board eeprom id’s
‘stat’ = report operational state, safety relay and tstat states and heater output power

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C