Appendix A: Specifications
SpO2 Specifications
A-6 Service Manual
The Masimo SpO
2
parameter has been validated for low perfusion accuracy in bench top testing against a
Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02% and a %
transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or
minus, one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
Note: Because pulse oximeter equipment measurements are statistically distributed, only about two-
thirds of pulse oximeter equipment measurements can be expected to fall within ± Arms of the
value measured by a COOximeter.
Pulse Rate accuracy has been validated for the range of 25-240 bpm in bench top testing against a Biotek
Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses 68% of
the population.
A.6.4.2 Specifications
Warmer Expected Service Life The product is designed to meet a life span of 8 years. However,
with proper maintenance and repairs, the service life can be
extended as long as service parts are available.
Measurement Range SpO
2
: 1-100% SpO
2
Pulse Rate: 25-240 bpm
Wavelengths
Red
Infrared
660 nm, nominal
905 nm, nominal
Maximum Optical Output Power
Radiant Power at 50 mA pulsed 0.13 mW, minimum
0.79 mW, maximum
Accuracy SpO
2
: from 70-100% SpO
2
, +/- 3% SpO
2
for neonates
Below 69%, unspecified
Pulse rate, no motion: +/- 3 bpm
Pulse rate, motion: +/- 5 bpm
Data Update Period At maximum averaging time, the data is updated in less than
thirty seconds; shorter averaging times reduce this update
period.
Alarms User selectable upper and lower limits for SpO
2
and pulse rates
Standards Designed to ISO 80601-2-61 with the exclusion to clause
201.15.3.5.101.1 with regard to shock and vibration.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C