EasyManua.ls Logo

GE PANDA

GE PANDA
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix D: Troubleshooting
Service Mode (Application Software 1.0.9 and Earlier)
D-6 Service Manual
“SILENCE” Byte 0 – Indicates the alarm that has been silenced.
This byte can be cross-referenced to Appendix C by adding 1.
IF Byte 0 = 9 then this indicated Alarm 10 on the table Temp. Probe Failure.
Byte 1 – The stage of the alarm. A “1” indicates the alarm has just occurred. A “2” indicates the alarm has
progressed because the alarm condition has not been alleviated or the alarm has not been silenced.
“SYSFAIL” Bytes 0-3 – The auxiliary data associated with the system failure.
Byte 4 – The system failure number which corresponds to the values found in the tables in section 4.4.5.
“STATE” Bytes 0-3 – The patient set point listed in hexadecimal byte-swapped format.
Note: No decimal place is present (6d 01 00 00 can be looked at as 0x0000016d which translates to 365 or
36.5 degrees C).
Byte 4 – Is used internally to determine what menu is on the screen.
Byte 5 – Is an indication of the mode.
“0x11” for Manual Mode.
“0x12” for Baby Mode.
“0x13” for Preheat Mode.
Byte 6 – Is an indication of the SpO
2
State.
“0” is not present.
“1” is present, but never attached to a patient.
“2” is no probe connected.
“3” is running.
“4” is stopped.
“5” is out of service.
Byte 7 – Is a hexadecimal indication of the percent power (from 0% to 100%).
“CFG” Byte 0 – Temp Units (0 for Centigrade, 1 for Fahrenheit).
Byte 1 – Alarm Volume.
Byte 2 – Elevating Base (0 for disabled, 1 for enabled).
Byte 3 – Hands Free Silence (0 for disabled, 1 for enabled).
Byte 4 – Patient Temp Alarm Setting.

Table of Contents

Related product manuals