EasyManuals Logo

GE PANDA Service Manual

GE PANDA
212 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #194 background imageLoading...
Page #194 background image
Appendix D: Troubleshooting
Service Mode (Application Software 1.0.9 and Earlier)
D-6 Service Manual
“SILENCE” Byte 0 – Indicates the alarm that has been silenced.
This byte can be cross-referenced to Appendix C by adding 1.
IF Byte 0 = 9 then this indicated Alarm 10 on the table Temp. Probe Failure.
Byte 1 – The stage of the alarm. A “1” indicates the alarm has just occurred. A “2” indicates the alarm has
progressed because the alarm condition has not been alleviated or the alarm has not been silenced.
“SYSFAIL” Bytes 0-3 – The auxiliary data associated with the system failure.
Byte 4 – The system failure number which corresponds to the values found in the tables in section 4.4.5.
“STATE” Bytes 0-3 – The patient set point listed in hexadecimal byte-swapped format.
Note: No decimal place is present (6d 01 00 00 can be looked at as 0x0000016d which translates to 365 or
36.5 degrees C).
Byte 4 – Is used internally to determine what menu is on the screen.
Byte 5 – Is an indication of the mode.
“0x11” for Manual Mode.
“0x12” for Baby Mode.
“0x13” for Preheat Mode.
Byte 6 – Is an indication of the SpO
2
State.
“0” is not present.
“1” is present, but never attached to a patient.
“2” is no probe connected.
“3” is running.
“4” is stopped.
“5” is out of service.
Byte 7 – Is a hexadecimal indication of the percent power (from 0% to 100%).
“CFG” Byte 0 – Temp Units (0 for Centigrade, 1 for Fahrenheit).
Byte 1 – Alarm Volume.
Byte 2 – Elevating Base (0 for disabled, 1 for enabled).
Byte 3 – Hands Free Silence (0 for disabled, 1 for enabled).
Byte 4 – Patient Temp Alarm Setting.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE PANDA and is the answer not in the manual?

GE PANDA Specifications

General IconGeneral
TypeInfant Warmer
ManufacturerGE Healthcare
Control ModesManual, Servo
Integrated ScaleYes
SpO2 MonitoringOptional
Bed TiltYes
Heater TypeCeramic
AlarmsPower Failure
Power Requirements100-240 VAC, 50/60 Hz
ModelPANDA
Resuscitation OptionsOptional
Skin Temperature Probe Accuracy±0.3°C

Related product manuals