5.5.2 Gas measurements
CO
2
O
2
(%)
N
2
O (%)
ISO
MAC
(%)
(%)
Fi
30
0.5
58 58
1.2
2.0
40 35
0.1
2.
0
5.3
6.5
30
4.0
2.4
2.4
Et.
Age: 40
The following values are monitored with the
patient gas analyzer:
Fi/Et O
2
Fi/Et CO
2
Fi/Et N
2
O
Fi/Et Agent
MAC
A reference measurement is made once every
four hours or if there is a shift in temperature
greater than 1
o
C. The message 'Gas analyzer
zeroing' will be displayed. During this
calibration period, the color of the gas
measurement values will turn gray, and the
last measured values will remain visible on the
screen. This reference measurement takes
approximately five seconds.
The same behaviour is observed in the case
of occlusion in the gas sampling line.
5.5.3 Gas supply pressure
312
01-01 13 00
O
2
Air N
2
O
4.3 3.9 4.5
4.8 3.5 4.3
The current gas supply pressure (mains supply
and backup gas supply) can be displayed in
the upper left area of the control panel. See
'Startup configuration', Chapter 9, page 174.
If N
2
O is not activated in the startup
configuration, 'N
2
O' and associated pressure
values are not shown.
If no backup gas is connected to the system,
the backup gas supply pressure values are
replaced with asterisks; '***'.
Current gas pressure can also be monitored
separately by activating the 'Gas pressure
tab', see page 112.
106
FLOW-i 4.2, User's Manual
| 5
| System functionality |

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge Maquet FLOW-i 4.2 and is the answer not in the manual?

Getinge Maquet FLOW-i 4.2 Specifications

General IconGeneral
BrandGetinge
ModelMaquet FLOW-i 4.2
CategoryMedical Equipment
LanguageEnglish

Related product manuals