9.4.3 Low priority clinical alarms
Color = black text, blue background.
Sound = A two tone pitch followed by a long
pause, repeated indefinitely.
The following low priority alarms exist on the
system:
DescriptionAlarm
Measured etO
2
concentration is above alarm limitEtO
2
: High
Air/N
2
O or O
2
central gas supply pressure <2.5 barAir/N
2
O/O
2
backup gas supply active
A mixture of anesthetic agents is detected. The
MAC of the secondary agent is 0.6, and the total
MAC is <3.
Agent mixture
Mains power failure, battery in operation.Battery operation
Liquid level <10%.Vaporizer liquid level low
200
FLOW-i 4.2, User's Manual
| 9
| Alarms and patient safety |