135
_____________________________________________________________________
7.5 Scaling Function
______________________________________________________________
Before scaling
After scaling
0
250
[
mV
]
-250
12.5
0
[
A
]
200 10
-12.5
NOTE
Refer to Section 7.3.4 to measure current using the HIOKI 9018-
10, 9132-10 CLAMP ON PROBEs.
(2) To scale a current waveform for direct reading from a
general purpose clamp sensor set to 10A current range that
provides 0.2V AC full scale output.
Either set a conversion ratio of 10:0.2 (50:1), or specify two
points as 0.2,10 and 0,0. The units are amperes [A].
Scaling set by conversion ratio
Scaling set by two points
When set as above, scaling is applied as shown below. The
voltage axis range is 50 mV/DIV.