134
_____________________________________________________________________
7.5 Scaling Function
______________________________________________________________
Before scaling After scaling
0
100
[
V
]
-100
10000
-10000
0
[
V
]
Scaling Examples
(1) To measure the voltage waveform at the secondary of a
power transformer with 100:1 winding ratio, the scaling
setting allows reading the measurement directly:
Either set a conversion ratio of 100,
or specify 2 points as 1 -> 100 and 0 -> 0.
The units are volts: [V]
Scaling set by conversion ratio
Scaling set by two points
When set as above, scaling is applied as shown below. The
voltage axis range is 20 V/DIV.